Introduction & Background: The gold standard method for measuring cerebral blood flow (CBF) is dynamic 15O-water PET with an arterial input function (AIF). The invasive nature of obtaining the AIF through arterial cannulation has hampered the widespread clinical use of the technique. Bach et al. (2025) proposed a dual scan protocol with an additional dynamic cardiac 15O-water PET scan for extracting an image-derived input function (IDIF) to replace the blood sampled AIF. However, this method increases patient dose significantly. In this study, we compare CBF quantification using IDIFs from retroactively subsampled cardiac scans.
Methods: Fourteen healthy controls each underwent dual-bolus 400 MBq 15O-water PET, one cardiac and one brain scan, six minutes each on a Discovery MI PET/CT (GE Healthcare). The cardiac scans were reconstructed using 100% (original), 50%, 25% and 10% of the recorded counts. IDIFs were extracted from the cardiac scans using two different masks. First, an arterial mask automatically derived from the original scan using the aQaunt software (Medtrace Pharma), and second, aQuant was used to automatically derive an arterial mask from each subsampled scan individually. Whole brain CBF was calculated using the single tissue compartment model with each of the seven different IDIFs.
Results: There was a strong agreement between CBF calculated with the IDIF from the original cardiac scan and the subsampled IDIFs, both when using identical arterial masks and individual masks automatically defined on each subsampled scan (Figure 1). Comparing 100% and 10% count reconstructions using the same arterial mask resulted in R2 = 0.997, slope = 0.99 CI: [0.96 1.02], and using the individual arterial masks in R2 = 0.986, slope = 0.96 CI: [0.89 1.03].
Conclusions: A cardiac PET scan with as little as 40 MBq can be used to accurately measure the arterial input function for 15O-water CBF quantification.